Air sample volume calculator

Flow rate × duration gives the liters of air drawn through a spore-trap cassette. Match indoor and outdoor volumes so the only difference your lab reads is the air itself.

L/min

Calibrated flow rate of your air-sampling pump.

min

How long the cassette ran.

Volume sampled75 L
In cubic meters0.075 m³
Within the typical spore-trap range for a representative residential air sample.

How air-sample volume works

A spore-trap result is only as good as the volume behind it. The sampled volume is simply the pump's calibrated flow rate (liters/minute) multiplied by the run time (minutes); dividing by 1,000 converts liters to cubic meters, the unit the lab reports counts in (spores/m³). A common protocol runs roughly 15 L/min for about 5 minutes — near 75 liters — which many labs treat as a practical floor for a representative residential air sample. There is no standard-published minimum volume; this is a guidance flag, not a pass/fail.

The volume that matters most is the match between your indoor and outdoor samples. An air result is read against a same-day outdoor control, so if the indoor cassette ran 5 minutes and the outdoor ran 3, you are comparing apples to oranges. Run identical flow and duration on both — see air pump calibration for why a calibrated, matched volume is the foundation of the indoor:outdoor comparison.

Related reading

Sources

Write the report in minutes, not hours.

MoldMind turns your field notes, photos, and lab results into a standards-compliant report you review and approve. Try MoldMind free — 3 jobs, no card.